Remote Airborne Particle Counters 9301 and 9501

Article: 9301

Article: 9501

The 9301 and 9501 particle counters provide a range of particle size measurement from 0.3 to 10.0 μm at a flow rate of 0.1 CFM (2.83 LPM).

GET A QUOTE

Description

The 9301 and 9501 particle counters provide a range of particle size measurement from 0.3 to 10.0 μm at a flow rate of 0.1 CFM (2.83 LPM). These sensors allow for efficient particle counting with specifications that exceed the ISO 21501-4 and JIS B9921 calibration standards.
These sensors have 6 user-selectable particle size channels (up to 30 channels optional) and are equipped with an integral critical orifice and flow sensor to provide a stable flow rate from an external vacuum source. Integration into a building automation or facility monitoring system is easy via isolated RS-485 Modbus communication.

Datasheet 9301/9501

Features and Benefits

• 9301: Measures 0.3 µm to 25 µm
• 9501: Measures 0.5 µm to 25 µm
• 0.1 CFM (2.83 LPM) flow rate
• Long life laser diode technology
• Measures up to 30 channels of simultaneous data
• Particle concentrations up to 10,000,000 Particles/ft3 @ 10% coincidence loss
• Internal critical orifice
• User-selectable channel sizes
• Stores up to 65,000 sample records for on-board data redundancy
• (Optional) temperature and relative humidity probe available
• Connect using Modbus RTU/ASCII over isolated RS-485
• Complies with ISO 21501-4 and JIS B9921 standards
• Easy to clean and wipe down with minimal particle traps
• Versatile mounting options
• Alarm light
• LED Indicators
• Seamless integration into a facility monitoring system
• Lightweight stainless steel enclosure
• 2 year warranty. Extended warranties available.

Specifications

Model 9301 9501
Size Range 0.3 µm to 25 µm 0.5 µm to 25 µm
Size Channels Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 µm Factory calibrated at 0.5, 0.7, 1.0, 2.5, 5.0, 10.0 µm
Number of Channels 6 channels (up to 30 optional) 6 channels (up to 30 optional)
Counting Efficiency 50% @ 0.3µm; 100% for particles > 0.45 µm per JIS 50% @ 0.5µm; 100% for particles > 0.75 µm per JIS
Flow Rate 0.1 CFM (2.83 LPM)
Concentration Limits 10,000,000 particles/ft3 @ 10% coincidence (per ISO 21501-4),

20,000,000 particles/ft3 @ 10% coincidence (as tested and validated1)

Light Source Long life laser diode
Zero Count < 1 count / 60 minutes (< 1 particles /6 ft3)
Alarms Channel alarms on Raw counts, concentrations or mass (alarms on environmental sensors optional)
Calibration NIST traceable
Vacuum Requirements External vacuum > 15” (38.1cm) of Hg
Airflow Internally monitored
Configuration/Download USB mini-B
Alarm Alarm LED ring
Communication Modes MODBUS™ RTU or ASCII outputs (over isolated RS-485)
Environmental Sensor (Optional) Temp and RH probe 32º to 122ºF (0º to 50ºC) ±1ºF (0.5ºC), 15-90% ±2% relative humidity
Standards ISO 21501-4 and JIS B9921
Instrument Calibration Recommended minimum once per year
External Surface Stainless steel
Dimensions (L x W x H) 3.59” x 1.83” x 5.01” (9.1 cm x 4.6 cm x 12.7 cm) including probes and connectors
Weight 1.08 lb. (494 grams)
Accessories Operating manual on USB flash drive, isokinetic probe, power supply and cable
Optional Accessories Printed manual barb fittings, mounting bracket and sample tubing, IMS-RT monitoring system
Buffer Memory 65,000 sample records (rotating buffer) including particle count data and environmental data
Sample Time 1 second to 99 hours
Power 9 – 24 VDC (< 1.5 watts)
Operating Conditions 41° to 104°F (5° to 40°C) / 20% to 95% non-condensing
Storage Conditions 32° to 122°F (0° to 50°C) / Up to 98% non-condensing
Warranty  2 year limited warranty. Extended warranties available.